Brand Name : Dräger
Model Number : 8422711
MOQ : 1pcs
Payment Terms : T/T(wire transfer), Western Union, Paypal
Delivery Time : 3-5 working days
Packaging Details : Carton
Product Name : PN 8422711 Dräger Savina Ventilator Power Supply EOE12060720
Brand : Dräger
Part No : 8422711
Input : 100-240V~/3,4-1.3A;50-60Hz
Ext : 24V-/13.5A
Int : 24V-/11.5A
Output : 5V-/4.5A;15V-/1.5A;-15V-/1.3A 24V-/1.2A;48V-/2.4A
Charge : 0-32V-/45W (27V-/1.66A)
Warranty : 90 days
MOQ : One piece
PN 8422711 Dräger Savina Ventilator Power Supply EOE12060720
Power supply
Savina runs on mains power or an internal battery. Savina can be powered by an external battery or built-in DC power supply. An external battery
is also used as a power source during patient transport.
Savina is designed for connection to the hospital's mains power supply.
WARNING
Risk of electric shock and device malfunction If the device is connected to a socket with the wrong mains voltage or without a protective earth
connection, the user is at risk and the device could be damaged. protective earth connector
NOTE
In operation, the power socket used must be readily accessible.
Power supply from on-board power supply or batteries
WARNING
Risk of explosion Electrolytic gas can occur when the batteries are charging. In a sufficient concentration, this can cause an explosion. The device
must always be placed in a wellventilated area when connected to mains power.
Supply from a DC on-board power supply
Savina can be powered by a DC on-board power supply
Connecting the power supply
Prerequisites: – Installation and connection by service personnel only. – The poles of the DC vehicle power supply must be protected against
contact
Warn
Risk of patient injury If the poles of the DC vehicle power supply are not protected against contact, excessive patient leakage currents may result.
Prevent poles from touching.
Warn
RISK OF EQUIPMENT DAMAGE - Do not connect mains powered equipment to the external battery connector. – Only external batteries included
in the accessories list can be connected. – Only use the connecting cables listed in the accessories list.
The plug on the device side of the connecting cable is coded. This allows Savina to detect if an external battery or DC vehicle power supply is connected.
Specification
Input | 100-240V~/3,4-1.3A;50-60Hz |
Ext | 24V-/13.5A |
Int | 24V-/11.5A |
Output | 5V-/4.5A;15V-/1.5A;-15V-/1.3A 24V-/1.2A;48V-/2.4A |
Charge | 0-32V-/45W (27V-/1.66A) |
Failure of the power supply
If the power supply fails, Savina generates a power supply failure alarm. The ventilation settings and the alarm limits remain saved even in the
event of a power supply failure